dc.contributor.author | Altuntaş, Derya Bal | |
dc.date.accessioned | 2020-12-19T19:35:00Z | |
dc.date.available | 2020-12-19T19:35:00Z | |
dc.date.issued | 2020 | |
dc.identifier.citation | Altuntaş, D.B. (2020). Development of All-Solid-State Antidiabetic Drug Metformin-Selective Microsensor and Its Electrochemical Applications. Electroanalysis, 32(6), 1280-1287. https://doi.org/10.1002/elan.201900773 | en_US |
dc.identifier.issn | 1040-0397 | |
dc.identifier.issn | 1521-4109 | |
dc.identifier.uri | https://doi.org/10.1002/elan.201900773 | |
dc.identifier.uri | https://hdl.handle.net/11436/1212 | |
dc.description | WOS: 000514639500001 | en_US |
dc.description.abstract | In this study, all-solid-state type potentiometric PVC membrane selective microsensor was developed for Metformin (MET) which is an antidiabetic drug active substance. Metformin-tetraphenylborate (MET-TPB) ion-pair was used as an ionophore in the structure of the sensor membrane. It was determined that the sensor membrane at the ratio of 69 % o-nitrophenyl octyl ether, 27 % polyvinyl chloride and 4 % MET-TPB performed the best potentiometric performance. in a wide concentration range (1x10(-5)-1x10(-1) mol/L), the slope, detection limit, response time, pH range, and life-time of the sensor were determined as 55.9 +/- 1.6 mV (R-2=0.996), 3.35x10(-6) mol/L, 8-10 s, pH: 3-8, and similar to 10 weeks, respectively. the voltammetric performances of the sensor were also investigated. the prepared microsensor was successfully utilized for the determination of Metformin in a pharmaceutical drug sample by potentiometry and voltammetry. It was observed that the obtained results were in agreement with the results obtained by the UV spectroscopy method at 95 % confidence level. | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Wiley-V C H Verlag Gmbh | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Antidiabetic | en_US |
dc.subject | Metformin | en_US |
dc.subject | Microsensor | en_US |
dc.subject | Potentiometry | en_US |
dc.subject | Voltammetry | en_US |
dc.title | Development of all-solid-state antidiabetic drug metformin-selective microsensor and its electrochemical applications | en_US |
dc.type | article | en_US |
dc.contributor.department | RTEÜ, Mühendislik ve Mimarlık Fakültesi, Biyomühendislik Bölümü | en_US |
dc.contributor.institutionauthor | Altuntaş, Derya Bal | |
dc.identifier.doi | 10.1002/elan.201900773 | |
dc.identifier.volume | 32 | en_US |
dc.identifier.issue | 6 | en_US |
dc.identifier.startpage | 1280 | en_US |
dc.identifier.endpage | 1287 | en_US |
dc.relation.journal | Electroanalysis | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |